G. Grinstein, R.H. Koch
Physical Review E - SNSMP
Measurements of TlBaCaCuO-based dc superconducting quantum interference devices (SQUIDs) performed using a practical feedback circuit gave a flat frequency response and an energy sensitivity of 6×10-30 J/Hz at 5 kHz and 77 K with an 80 pH SQUID. Flux-locked loop operation was demonstrated to be strongly effective in eliminating the magnetic field hysteresis of the SQUID. The SQUIDs were operated in the flux-locked loop at temperatures over 90 K and the 1/f noise was found to decrease with increasing temperature near 77 K.
G. Grinstein, R.H. Koch
Physical Review E - SNSMP
R. Beyers, G. Lim, et al.
Applied Physics Letters
W.M. Goubau, P.M. Maxton, et al.
Geophysics
Y. Lu, P.L. Trouilloud, et al.